Autofocus Tube

Radiology & Physiotherapy, Chemistry & Physics

Also known as: Automatic Focus X-ray Tube, Variable Focus X-ray Tube

Historical document, translated for reference. It reflects medical knowledge of the 1920s–30s and is not medical advice.

Summary

An autofocus tube is an X-ray tube where the focal spot area automatically changes according to the tube's load. This is achieved through a device that adjusts the electron beam dispersion based on current flow, protecting the anticathode from rapid destruction at high loads.

Encyclopedia article (1928–1936)

Autofocus tube, an X-ray tube, the focal spot area of which is automatically changed in accordance with changes in the tube load. The automatic change in focal spot area is achieved with the help of a device, schematically shown in the figure. In the middle of the turns of an ordinary flat spiral cathode, perpendicular to its plane, a rod S is insulated from the turns. The high voltage from the transformer is applied directly to rod S, while to the spiral turns through a large resistance R. During tube operation, the rod will always have a slightly greater negative potential relative to the spiral, and this potential difference must depend on the strength of the current flowing through the tube. Let resistance R = 10,000 ohms; then, with a tube current of 100 mA, the potential drop in resistance R will be (according to Ohm's law) 1,000 volts. By this amount, the potential of the rod will be higher than the potential of the spiral. With a tube current of 1 mA, the difference between the negative potentials of the rod and the spiral will only be 10 volts. Rod S, having a more negative potential than the spiral, will always influence the path of the electrons from the spiral such that they will be somewhat scattered in their movement toward the anticathode, forming a focus with a larger area. This increase in focal spot area will be greater the more the potential of the rod exceeds the potential of the spiral, i.e., it will be, as indicated above, at higher tube loads. The increase in focal spot size at higher loads protects the tube's anticathode from rapid destruction, since the energy lost by the electrons is distributed over a larger surface.

v. Karchagin.

Autofocus Tube: figure 1 from the 1928–1936 encyclopedia article

Cite this page

“Autofocus Tube.” Soviet Medical Encyclopedia. English translation of Bolshaya Meditsinskaya Entsiklopediya, 1st ed. (Moscow, 1928–1936), ed. N. A. Semashko. https://sovietmedicalencyclopedia.pages.dev/article/autofocus-tube/